Error Messages - AIR160

The software stops if there is a problem with the real time clock or with the ventilator connection. As soon as the device is switched on, the program will carry out a self-test to check the proper operation of the product. The units that are checked are the real time clock and the ventilator connection. Any problems with these items will be indicated on the display.

This message indicates that there is a problem with the BU battery. The BU battery normally has a lifespan of 10 years. In the unlikely event that this is different, this message will appear.

If you get this message please contact your local dealer to find a solution.

FAIL002 is a warning shown in the display of the unit if the unit has been switched off. The Air 160 needs to be switched on 24/7, this to make sure that it can offer the most reliable disinfection of your room.If the unit is switched off the warning message FAIL002 will appear in the display.

This warning message indicates that the room is not disinfected and may hold some unwanted micro-organisms.

Pressing the button on the display of the AIR 160 once will simply remove this warning message.

For more info?

Do you have a question about our products and would you like to contact us? Please call us or complete the form below.

We will be happy to help.

Display messages - AIR160

If you press the button once in the rest position, the operating hours will be displayed. The display will automatically start from the factory setting of the timer, i.e. 8,000 hours. 

The UV-C lamp operates at 100% radiation for up to 8,000 hours. After 8,000 operating hours, the UV-C radiation will weaken, and may be insufficient to obtain the desired results. If the device is switched off and then on again, the hour counter will start counting backwards from the value it had when it was switched off.

The display will indicate 0000 when the 8,000 operating hours of the lamp have expired. The actual time will blink in the display, alternating with four blinking zeros. The filter and the UV-C lamp must now be replaced. See item ‘rSt’ in your manual “How to reset the timer”.